SciELO - Scientific Electronic Library Online

 
vol.26 issue3Aerobic exercise in patients with heart failure with and without ventricular dysfunction in a cardiac rehabilitation programme author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

  • On index processCited by Google
  • Have no similar articlesSimilars in SciELO
  • On index processSimilars in Google

Share


Revista Colombiana de Cardiología

Print version ISSN 0120-5633

Abstract

SEGUEL, Enrique et al. Experimental implant of a continuous flow circulation pump in an aortic valve position. Rev. Colomb. Cardiol. [online]. 2019, vol.26, n.3, pp.169-176. ISSN 0120-5633.  https://doi.org/10.1016/j.rccar.2018.08.009.

Introduction:

The implanting of an axial flow pump in an aortic valve position could decrease certain complications in current pumps. It could permit an anterograde blood flow in the heart and the aorta. It could avoid the making of a bag, as well as avoiding the need for inlet and outlet cannulas, and could slow down the scar formation in the cardiac apex.

Objective:

To study the technical feasibility of the implanting of an axial flow pump in an aortic valve position, as well as to assess the haemodynamic effects and its impact on cardiac structures.

Methods:

A Jarvik 2000® pump adapted for an intra-valvular implant in 7 pigs was inserted using left anterior thoracotomy and under extracorporeal circulation. The pump was fixed to the aortic ring using sutures. It assisted the subjects for an hour during an assessment was made of the haemodynamic behaviour and the cardiac structures with an epicardial echocardiogram. After euthanasia of the animal, the heart was removed for an examination of its cavities.

Results:

All the procedures were completed with the technique. With the vasoactive drugs and epicardial pacemaker support, the pump was capable of maintaining the haemodynamics of the individuals. The echocardiogram showed adequate filling of the left ventricle and good functioning of the mitral valve. The post-mortem examination showed that the coronary ostia were free and there had not been any clots in the cardiac cavities.

Conclusions:

In this animal model, the implanting of an axial flow pump in an aortic valve position was possible. The pump was able to maintain the haemodynamics of the animal without significantly affecting the cardiac structures.

Keywords : Cardiac assist device; Animal models; Heart failure; Cardiac surgery; Bio-implants; Jarvik 2000..

        · abstract in Spanish     · text in Spanish     · Spanish ( pdf )