SciELO - Scientific Electronic Library Online

 
vol.51 número2Efectos adversos severos de los estabilizadores del ánimo: revisión de la literaturaAnálisis de la evolución del conocimiento preventivo y su práctica entre la población de Colombia hacia la enfermedad por Coronavirus (Covid-19) índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • No hay articulos similaresSimilares en SciELO
  • En proceso de indezaciónSimilares en Google

Compartir


Revista Colombiana de Ciencias Químico - Farmacéuticas

versión impresa ISSN 0034-7418versión On-line ISSN 1909-6356

Resumen

GALANA GERLIN, Mirella Carla et al. Development and validation of an analytical method by HPLC for the simultaneous determination of atorvastatin, losartan and metformin in compounded pharmaceutical formulations. Rev. colomb. cienc. quim. farm. [online]. 2022, vol.51, n.2, pp.955-970.  Epub 21-Dic-2023. ISSN 0034-7418.  https://doi.org/10.15446/rcciquifa.v51n2.96964.

Aims:

To develop and validate an analytical method for the simultaneous analysis of atorvastatin (ATO), losartan (LOS), and metformin (MTF) in pharmaceutical compounding by HPLC.

Method:

The method was developed using a liquid chromatograph model Dionex® Ultimate 3000, coupled to a diode array detector (DAD), and validation was carried out according to international and national validation compendiums. The following parameters were analyzed: linearity, precision, accuracy, robustness, selectivity, LD, and LQ. Chromatographic separation was performed on a Sigma-Aldrich® C18 column, mobile phase MeOH:H2O (78:22 v/v), pH 3.0, and flow rate of 0.3 mL-min-1, isocratic mode. Detections were performed at 225 nm (LOS), 236 nm (MTF), and 246 nm (ATO) wavelengths.

Results:

The method was linear, with linear correlation coefficients (r)> 0.99 for all drugs, precise (DPR between measurements < 2 %), accurate recovery (between 98-102 %). The variations proposed in the robustness study had no significant influence on the results, except for the chromatographic column. The method was selective, as the excipients did not interfere in the analyses. The sensitivity of the method was demonstrated through the theoretical determination of LD and LQ.

Conclusion:

The results obtained suggest that the chromatographic method developed and validated in this research can be used in the routine analyzes of quality control laboratories of drugs containing these drugs alone or combined in pharmaceutical formulations for oral use.

Palabras clave : Antihypertensive; antidiabetics; antilipemic, quality control; HPLC.

        · resumen en Portugués     · texto en Portugués     · Portugués ( pdf )