vol.65 número2Práctica profesional de los egresados del programa de Fisioterapia de la Universidad Nacional de Colombia en el marco del sistema de salud colombianoEntrevista: guía práctica para la recolección de datos cualitativos en investigación de salud índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • No hay articulos similaresSimilares en SciELO
  • En proceso de indezaciónSimilares en Google

Compartir


Revista de la Facultad de Medicina

versión impresa ISSN 0120-0011

Resumen

MUNOZ, Isabel Cristina  y  HERNANDEZ, Alher Mauricio. Changes in ventilatory mechanics caused by variations in PEEP and pressure support: study in healthy subjects under non-invasive mechanical ventilation. rev.fac.med. [online]. 2017, vol.65, n.2, pp.321-328. ISSN 0120-0011.  https://doi.org/10.15446/revfacmed.v65n2.60938.

Introduction:

Traditionally, ventilatory mechanics has been delivered in controlled modes through the use of inappropriate approaches for spontaneous ventilation.

Objective:

To measure the changes of ventilatory mechanics caused by PEEP and pressure support (PS) variations in non-invasive mechanical ventilation.

Materials and methods:

The ventilatory mechanics was evaluated through a non-invasive strategy, under different PEEP and pressure support levels. For this purpose, a mechanical simulator was used, and a database of 14 healthy subjects non-invasively connected to a mechanical ventilator was recorded.

Results:

For the 14 healthy subjects under PEEP and pressure support conditions of 0 cmH2O, the median resistance and compliance values were 91.2 [77.8-135.9] mL/cmH2O and 8.3[6.1-10.4]cmH2O/L/s, respectively. PEEP compliance showed a statistically significant increase in all subjects. On the other hand, no changes in any of the parameters were observed regarding pressure support increase.

Conclusions:

The proposed technique allowed to find compliance and resistance values consistent with those set in the mechanical simulator, which, in turn, coincide with those reported in the literature for healthy subjects. This information is useful for decision-making in intensive care units.

Palabras clave : Noninvasive Ventilation; Respiratory Mechanics; Work of Breathing (MeSH).

        · resumen en Español     · texto en Español     · Español ( pdf )