SciELO - Scientific Electronic Library Online

 
vol.47 issue2Evaluation of a closed-loop intravenous total anesthesia delivery system with BIS monitoring compared to an open-loop target-controlled infusion (TCI) system: randomized controlled clinical trialManagement and perioperative outcomes of traumatic brain injury: retrospective study author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

  • On index processCited by Google
  • Have no similar articlesSimilars in SciELO
  • On index processSimilars in Google

Share


Colombian Journal of Anestesiology

Print version ISSN 0120-3347

Abstract

PRADA-MANCILLA, William Andrés et al. Comparison of effectiveness between a conventional catheter and a manual commercial catheter: randomized trial of simulation of intraosseous access in a biological model. Rev. colomb. anestesiol. [online]. 2019, vol.47, n.2, pp.92-99.  Epub May 01, 2019. ISSN 0120-3347.  https://doi.org/10.1097/cj9.0000000000000101.

Introduction:

In the emergency services an action of paramount importance in critically ill patients is to obtain an early vascular access. When vascular access is not established, an intraosseous route should be obtained; otherwise, the mortality of these critically ill patients is almost 100%. In Colombia, the intraosseous access is not used because of the high costs of the devices and the lack of training of the healthcare staff to conduct the procedure.

Objective:

To determine the efficacy of a low-cost device to establish intraosseous access.

Materials and methods:

Quantitative approach, experimental design in a simulated environment with chicken tarsus and metatarsus. An analysis was conducted using frequency tables and central tendency measurements. Likewise, further analyses were done using Fisher's exact test, Chi2, and Mann-Whitney test.

Results:

A total of 99% of the procedures were successful with both catheters. The average time for intraosseous access was 6.6 seconds with Insyte 14 catheter and 4.7 seconds with Din 1515x Illinois Desch device (P = 0.001). There were no significant differences in the number of attempts to secure a successful intra-osseous access using any of the 2 devices (P = 0.56).

Conclusion:

There was no significant difference between the Ci 14 and the Si 14 catheter to establish a successful intraosseous access in the chicken tarsus and metatarsus in a simulated environment.

Keywords : Infusions; Intraosseous; Clinical Trials as Topic; Vascular Access Devices; Critical Care; Efficacy.

        · abstract in Spanish     · text in English | Spanish     · English ( pdf ) | Spanish ( pdf )