vol.29 número3Efectividad de la citología cérvico-uterina para la detección temprana de cáncer de cuello uterino en el marco del sistema de salud de ColombiaDetección de efectos secundarios asociados a la administración de tramadol y dipirona en un hospital de alta complejidad índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • No hay articulos similaresSimilares en SciELO
  • En proceso de indezaciónSimilares en Google

Compartir


Biomédica

versión impresa ISSN 0120-4157versión On-line ISSN 2590-7379

Resumen

NUNEZ, María Ximena  y  BLANCO, Claudia. Efficacy of Orbscan II® and Pentacam® topographers by a repeatability analysis when assessing elevation maps in candidates to refractive surgery. Biomédica [online]. 2009, vol.29, n.3, pp.362-368. ISSN 0120-4157.

Introduction: Anterior and posterior corneal elevations are measurements used to detect keratoconus suspects. Purpose: To determine the efficacy of Orbscan II® and Pentacam® when assessing their elevation maps. Materials and methods: The efficacy of the Orbscan II and Pentacam measuring the anterior and posterior corneal elevations were evaluated in a sample of 68 eyes. The concordance between the two devices and the coefficient of repeatability were measured following the parameters of the British Standard Institution by the Bland-Altman concordance analysis and the Lin concordance correlation coefficient. Results: The coefficient of repeatability at the point of maximum anterior elevation was 68.29% with the Orbscan and 24.20% with the Pentacam. The concordance correlation coefficient was 0.64 (CI 95%: 0.48-0.76) with the Orbscan and 0.94 with the Pentacam (CI 95%: 0.91-0.96). The coefficient of repeatability at the point of maximum posterior elevation was 38.7% with the Orbscan and 68.0% with the Pentacam. The concordance correlation coefficient was 0.69 with the Orbscan (CI 95%: 0.55-0.80) with a precision of 0.71 and an accuracy of 0.97, and 0.24 with the Pentacam (CI 95%: 0.00-0.45) with a precision of 0.24 and an accuracy of 0.99. Conclusions: Measurement of the point of maximum posterior elevation is better with the Orbscan II and less precise with the Pentacam. The random error can be reduced by using the mean of three assessments and can serve as a guide in the search of diagnostic devices with minimum absolute relative error in all measurements.

Palabras clave : cornea; corneal topography; keratoconus; reproducibility of results; selection bias; efficacy.