SciELO - Scientific Electronic Library Online

 
vol.34 número2Complications of the cochlear implant in adults and children in the Hospital Universitario Doctor Peset de Valencia, Spain, from 2001 to 2020 índice de autoresíndice de assuntospesquisa de artigos
Home Pagelista alfabética de periódicos  

Serviços Personalizados

Journal

Artigo

Indicadores

Links relacionados

  • Em processo de indexaçãoCitado por Google
  • Não possue artigos similaresSimilares em SciELO
  • Em processo de indexaçãoSimilares em Google

Compartilhar


Medicas UIS

versão impressa ISSN 0121-0319versão On-line ISSN 1794-5240

Resumo

SALCEDO-CIFUENTES, Mercedes et al. Performance evaluation of a POCT analyzer in relation to the results of routine conventional clinical laboratory methods in a nursing home. Cali. 2014. Medicas UIS [online]. 2021, vol.34, n.2, pp.9-18.  Epub 28-Set-2021. ISSN 0121-0319.  https://doi.org/10.18273/revmed.v34n2-2021001.

Introduction:

Point-of-care testing (POCT) are clinical diagnostic tests carried out than laboratory analysts, outside of spaces specifically designed for clinical analysis, and they provide quick results that improve the timeliness of medical decision making. In Colombia there is no information on its use and performance in specific age groups such as those in nursing homes.

Objective:

To evaluate the performance of a POCT analyzer for lipid profile (CT, LDL-c, HDL-c, TG) and glycemia in relation to the results of routine conventional methods of the clinical laboratory in a nursing home.

Materials and methods:

Descriptive cross-sectional. 52 residents were taken to whom paired samples were applied. Glucose and lipid levels were determined. Samples collected by fingerstick were analyzed by POCT and venipuncture by conventional methods certified by the CDC in the laboratory. A standardized instrument was used to describe the desired characteristics of the POCT. Univariate and bivariate statistics were applied. The results issued by the clinical reference laboratory were compared with those of the POCT through the ICC.

Results:

The average age of the participants was 78, range 64-91 years. The POCT showed an acceptable performance compared to conventional clinical laboratory methods, especially TG and HDL-c. However, statistically significant differences were observed in the results of glycemia, CT and LDL-c delivered by the POCT compared to those of the clinical laboratory.

Conclusions:

POCT can be an important option for chronic disease screening and management in nursing homes. However, an organizational structure is necessary to ensure the quality of the POCT measurements. MÉD.UIS.2021;34(2): 9-18.

Palavras-chave : Health Services for the Aged Point-of-care testing; Analytical chemistry techniques; Instrumentation.

        · resumo em Espanhol     · texto em Espanhol     · Espanhol ( pdf )