vol.38 número1Análisis del comportamiento tenso-deformacional de un suelo parcialmente saturado para un cimiento circular bajo carga axialDisminución de la opacidad en las emisiones de gases contaminantes en motores de encendido por compresión mediante el uso de biodiésel B5 y B10 a base de algas (chlorella) índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • No hay articulos similaresSimilares en SciELO
  • En proceso de indezaciónSimilares en Google

Compartir


Ingeniería y Desarrollo

versión impresa ISSN 0122-3461versión On-line ISSN 2145-9371

Resumen

LEAN, PAUL; ALVAREZ, CARLOS; NORIEGA, AINSWORTH  y  NAVARRO, ERIK. Development and validation of the mechanical behavior of a cervical disc implant with an "S" shape of Ti-6AI-4V ELI manufactured by SLM. Ing. Desarro. [online]. 2020, vol.38, n.1, pp.197-211.  Epub 11-Jun-2021. ISSN 0122-3461.  https://doi.org/10.14482/inde.38.1.620.5.

In the present study the mechanical behavior of a titanium alloy cervical medical implant type Ti-6Al-4V ELI is developed and validated, manufactured by Selective Laser Melting (SLM), which is a highly versatile and a faster process that allows to manufacture components with complex shapes. The medical implant developed in this research presents improvements compared to existing ones. The proposal of this research is the geometric design optimization set up as an "S" shape, by using finite elements analysis and the feedback of the results of mechanical tests. Chemical analyzes, microstructural analyzes, mechanical tests and finite element analysis were carried out, taking as reference the guidelines established in the ASTM F136-13 and ASTM 2346-11 standards. During the development of this project, several geometric modifications were presented as a result of the constant feedback between the simulation of the design and the behavior of the implant under cyclic loads. Finally, the optimal design of the implant was found and this design achieves the functional requirements of a cervical implant.

Palabras clave : ASTM F136; ASTM 2346; 3d fatigue behavior; cervical implant; 3d printing; finite element analysis in 3d printing; ELI alloy; Ti6AL4V.