vol.14 número1Validación del material lúdico de la estrategia educativa basada en juegos para la promoción de estilos de vida saludable en niños de cuatro a cinco años de edadRelación de la dieta con los niveles plasmáticos de lípidos y lipoproteínas en sujetos adultos con diferentes genotipos del gen de la apolipoproteína E índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • No hay articulos similaresSimilares en SciELO
  • En proceso de indezaciónSimilares en Google

Compartir


Universitas Scientiarum

versión impresa ISSN 0122-7483

Resumen

CASTRO, Isabel-Cristina et al. Elaboration of alanine dosimeters and their possible applications in professional risks. Univ. Sci. [online]. 2009, vol.14, n.1, pp.86-91. ISSN 0122-7483.

Objective. To develop dosimeters of alanine and alanine/paraffin in order to meet the growing use of ionizing radiation at the technological level. Materials and methods. Free radicals produced by ionizing radiation in alanine are detected by electron spin resonance (ESR); paraffin is used as binder. The methodology of preparation includes: elaboration, irradiation, data collection and data analysis for the two types of dosimeters. The ESR spectra studied correspond to alanine dosimeters with masses between 60 and 120 mg and irradiated with doses of 10 and 20 Gy, and alanine/paraffin dosimeters with 160 mg and irradiated with different doses in a beam of photons 4 MV. Results. The intensity of the spectra (the characteristic five lines) depends on the doses received by the dosimeters and their relationship is linear. With pure alanine and a dose of 10 Gy, the minimum amount required was 120 mg; the achieved cylinder compaction was not sufficient to avoid the partial fragmentation of the dosimeter. Paraffin has no paramagnetic signal; in the alanine/paraffin dosimeters (cylinders of 13 mm in length and 3.5 mm of diameter, with a ratio 80:20, and good hardness) the response signal ESR-dose was studied in a range between 20 and 120 Gy. Conclusions. The cylinders of alanine/paraffin manufactured have an appropriate hardness to be handled as dosimeters of ionizing radiation; their reproducibility and effectiveness in the accumulation of dose is good.

Palabras clave : dosimetry; ionizing radiation; spectrometry of electron spin resonance (ESR); alanine dosimeters; alanine/paraffin dosimeters.