vol.25 issue1Manipulation of a robotic endoscope holder with encephalographic signals in a surgical simulatorStrengthening communication channels for people with hearing and speech disabilities in basic education settings through motion capture using sign languages author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

  • On index processCited by Google
  • Have no similar articlesSimilars in SciELO
  • On index processSimilars in Google

Share


Ingeniería y competitividad

Print version ISSN 0123-3033On-line version ISSN 2027-8284

Abstract

WILCHES-VISBAL, Jorge H.  and  APAZA-VELIZ, Danny G.. Validation of the spectral reconstruction of an electron beam by comparison with spectrum from Monte Carlo. Ing. compet. [online]. 2023, vol.25, n.1, e-21112051.  Epub Dec 30, 2022. ISSN 0123-3033.  https://doi.org/10.25100/iyc.v25i1.12051.

Inverse reconstruction, one of the ways of calculating the energy spectrum of the central axis, has shown good results in various studies. Its validation, in the absence of the real spectrum and practicality, is usually done by comparing the PDP measured in the clinic, using the gamma index. The objective of the work is to validate the spectral reconstruction of a 6MeV electron beam from a Sinergy linear accelerator, by comparison with the spectrum derived from the Monte Carlo simulation of the accelerator head, as well as to carry out an analysis of its relationship with the validation by means of the clinically accepted gamma index criterion (>95% within 3% dose difference/3 mm in distance to agreement). The inverse reconstruction used Tikhonov regularization and generalized simulated annealing. Excellent agreement was observed between the PDP and the reconstructed dose profile (obtained from the reconstructed spectrum) and the measured and simulated PDPs (from the head simulation spectrum), as passage was >95% within of 1%/1mm and 2%/2mm for the PDPs and the dose profile, respectively. The simulated and reconstructed spectra presented similar shapes, coinciding in most probable energy and much of the low energy region. Despite serious discrepancies in the peak region, this was not reflected in clinically observable differences. In conclusion, it was verified that the reconstructed spectrum is close to one simulated by Monte Carlo, making it appropriate for clinical and research use.

Keywords : Depth dose percentage; dose profile; reverse reconstruction; Monte Carlo validation; gamma index.