vol.8 número16IDENTIFICACIÓN DE ZONAS SUSCEPTIBLES DE ABLACIÓN EN FIBRILACIÓN AURICULAR PERMANENTE IMPLEMENTANDO ENTROPÍA APROXIMADA EN UN MODELO 2DLOCALIZACIÓN DE FOCOS ECTÓPICOS DURANTE FIBRILACIÓN AURICULAR PERMANENTE, MEDIANTE UN NUEVO ÍNDICE OHF. ESTUDIO DE SIMULACIÓN índice de autoresíndice de materiabúsqueda de artículos
Home Pagelista alfabética de revistas  

Servicios Personalizados

Revista

Articulo

Indicadores

Links relacionados

  • En proceso de indezaciónCitado por Google
  • No hay articulos similaresSimilares en SciELO
  • En proceso de indezaciónSimilares en Google

Compartir


Revista Ingeniería Biomédica

versión impresa ISSN 1909-9762

Resumen

ROLDAN VALENCIA, Frank Marcel; ESCOBAR RESTREPO, Julio César; LATORRE CORREA, Federico  y  VILLARRAGA OSSA, Junes. LINEAR AND NONLINEAR ANALYSIS EFFORTS IN ABUTMENTS OF PREFORMED ZIRCONIUM. FINITE ELEMENT. Rev. ing. biomed. [online]. 2014, vol.8, n.16, pp.39-50. ISSN 1909-9762.

The purpose of this study was to determine the mechanical behavior between Zimmer® zirconium implant abutments with and without a titanium washer at their base, by using the finite element analysis (FEA). Materials and Methods: two models of a zirconium abutment on implants were designed, composed of a Zimmer® titanium implant a fixation screw, cortical and cancellous bone, and the ceramic crown of an upper central incisor. An oblique force was applied with a magnitude of 200N, ascending in a scale of 200N until any of the components of the model exhibited a nonlinear behavior. Von Mises stresses were calculated for each element. Results: No significant differences on von Mises stresses were observed between the abutments with and without a titanium washer (280, 1MPa with a titanium washer, and 280, 4 MPa without it) but differences were found at the fixation screw and the implant: the stresses were lower in the model with a titanium washer (474MPa vs 576,7MPa). Both the bone and the crown exhibited non linearity at 600N, the implant and the screw at 1000N, and both zirconium abutments at 1400N. Conclusions: When the preload stress applied to both models is the adequate, no mechanical behavior differences between the abutments occur.

Palabras clave : Mechanical behavior; Zirconium abutment; Finite element method; Non linear.