Case report

Histological artifacts associated with laser and electroscalpel gingivectomy: Case series

Artefactos histológicos asociados a gingivectomía con láser y electrobisturí: serie de casos

Jennifer Orozco1  * 

David Rico1 

Lía Barrios2 

Vivi Hoyos1 

Pilar Blanco3 

1 Programa de Odontología, Corporación Universitaria Rafael Núñez, Cartagena, Colombia

2 Grupo de Histopatología, Facultad de Medicina, Universidad de Cartagena, Cartagena, Colombia

3 Medical Laser Latinoamérica, Santa Marta, Colombia


Abstract

Introduction.

Over time, efforts have been invested in the design of new instruments that overcome the disadvantages of the gold standard instrument in surgery, the scalpel. As a result, electronic equipment has emerged such as the electric scalpel and laser devices. The available evidence on these instruments suggests that the tissue response is related to each instrument’s physical and biological cutting principles.

Objective.

To compare the histological changes in gingiva samples associated with surgical cutting performed with a 940 nm diode laser, a 2780 nm erbium, chromium: yttriumscandium-gallium-garnet (Er,Cr:YSGG) laser, and an electric scalpel, by presenting a series of cases.

Case presentation.

We present three cases of healthy patients undergoing cosmetic surgery. The clinical examination revealed exposure of a keratinized gingiva band greater than 4 mm, normal color and texture in gingival tissue, with a firm consistency and no bleeding on periodontal probing. Gingivectomy was indicated with the following protocols: Diode laser of 940 nm at 1 W, in continuous mode; Er,Cr:YSGG laser of 2780 nm at 2.5 W, 75 Hz, H mode, air 20, water 40, gold tip MT4); and electric scalpel in cutting mode at power level four. Gingival tissue samples were taken and stored in 10% formaldehyde for histological analysis.

Conclusion.

All the evaluated cutting instruments generated histological changes produced by the thermal effect, the main ones being collagen coagulation and carbonization. The depth of thermal damage caused by the 2780 nm Er,Cr:YSGG laser was much lesser than that induced by the electric scalpel and the 940 nm diode laser.

Keywords: Gingivectomy; artifacts; histology; lasers, semiconductor; lasers, solid-state

Resumen

Introducción.

Históricamente se ha invertido esfuerzo en el diseño de nuevos instrumentos que superen las desventajas del estándar de referencia en cirugía, el bisturí. Como consecuencia de esto, han surgido equipos electrónicos como el electrobisturí y los diferentes dispositivos de tecnología láser. La información disponible sobre estos instrumentos sugiere que la respuesta del tejido intervenido está influenciada por los principios físicos y biológicos de corte del instrumento.

Objetivo.

Comparar los cambios histológicos en muestras de encía asociados al corte quirúrgico realizado con láser de diodo de 940 nm, láser de erbio, cromo: itrio-escandio-galio-granate (Er,Cr:YSGG) (2780nm) y electrobisturí mediante una presentación de serie de casos.

Presentación de los casos.

Se presentan tres casos de pacientes sanos sometidos a cirugía estética. El examen clínico reveló la exposición de una banda gingival queratinizada mayor de 4 mm, tejido gingival de color y textura normales, de consistencia firme y sin sangrado al sondaje periodontal. Se indicó gingivectomía con los siguientes protocolos: láser de diodo de 940 nm a 1 W, en modo continuo; láser de Er,Cr:YSGG de 2780 nm a 2,5 W, 75 Hz, modo H, aire 20, agua 40, punta de oro MT4; y bisturí eléctrico en modo de corte, a nivel de potencia cuatro. Se tomaron muestras de tejido gingival y se almacenaron en formaldehído al 10 % para su análisis histológico.

Conclusión.

Los tres instrumentos de corte generaron cambios histológicos producidos por el efecto térmico; los principales fueron coagulación del colágeno y carbonización. La profundización del daño térmico causada por el láser de Er,Cr:YSGG de 2780 nm fue mucho menor que la generada por el electrobisturí y por el láser de diodo de 940 nm.

Palabras clave. Gingivectomía; artefactos; histología; láseres de semiconductores; láseres de estado sólido

Today, aesthetics is one of the most frequent reasons for patient’s dental visits 1. It is common to find cases of gummy smiles, where the patient manifests dissatisfaction due to the reduced size of the teeth and the amount of gingiva shown when smiling 2. If the primary cause of the gummy smile is periodontal, the conventional treatment seeks to increase the clinical crown of the teeth through a gingivectomy 3. This procedure is traditionally performed with a scalpel or electric scalpel. However, with the advent of new technologies in dentistry, laser technology could represent a surgical alternative with multiple benefits, and its application, for years, has been reported by different authors 4-6.

The 940 nm diode laser and the 2780 nm erbio, cromo: itrio-escandio-galio-granate (Er,Cr:YSGG) laser have entered the dental field with a wide range of applications in different areas 7,8. Its main benefits in soft tissue surgeries are reduction or absence of intraoperative bleeding, speeded healing, reduction of postoperative pain and inflammation, absence of risk of damaging the adjacent tooth or bone, excellent hemostasis, reduction of postoperative bacterial contamination, and therefore, reduced risk of infection, light contact of the fiber tip with the tissue, proprioceptive feedback and good patient acceptance, especially in those procedures requiring only local anesthesia 9,10.

A controlled clinical trial, carried out by Sobouti et al. in 30 patients who underwent gingivectomy to correct a gummy smile, showed that patients treated with the 940 nm diode laser had a significantly lower bleeding rate (0.36 out of 4) than that observed in the group operated with a conventional scalpel (1.15 out of 4). Likewise, it was observed that none of the diode laser participants showed post-surgical pain, nor did they consume analgesics, while in the control group, an average pain of 5.2 was obtained (out of a scale of 10), and 14 of the participants had to take pain medication 11.

This precedent shows the relevance of the instrument choice for soft tissue surgeries to provide comfort and results to the patient. Therefore, it is important to consider some aspects such as intraoperative and postoperative symptoms and time and mechanism of tissue healing, directly related to the effect exerted by the instrument to perform the cut.

From this point of view, an incision in the gingiva comprises numerous events depending on the instrument used, from damage to the cells and the extracellular matrix to the complete carbonization of the tissue, implying a greater difficulty in the wound repair and healing processes 12. Few studies report the histological effect on gingival tissue exerted by cutting instruments.

This article aims to compare the histological changes in gingiva associated with surgical cuts made with a 940 nm diode laser, a 2780 nm Er,Cr:YSGGE laser, and an electric scalpel through a series of cases.

Case presentation

Three clinical cases are presented: Two female patients and one male, who attended the dental clinic of the Corporación Universitaria Rafael Núñez in Cartagena due to aesthetic discomfort. All patients reported: “I display a large amount of gum when smiling”.

During the anamnesis, patients did not report systemic pathologies or relevant personal or family medical history. When performing the stomatological examination, an exposure of keratinized gingiva greater than 4 mm was observed in the three cases. The tissue was healthy, of normal color and texture, with a firm consistency and no bleeding on periodontal probing (figure 1).

Figure 1 Initial photos. The clinical characteristics of the periodontium of each patient are observed. 

As a treatment plan, gingivectomy and gingivoplasty were indicated in the upper jaw. Before the surgical procedure, oral health education was carried out to reinforce hygiene habits. The procedures were performed as follows:

Electrosurgical gingivectomy: Two percent lidocaine plus 1:80,000 diluted epinephrine was used as local anesthesia. Probing was performed with a Michigan periodontal probe to determine the height and amount of gingival tissue to be cut. The cutting instrument used was an ART® Electrosurgery Unit in cut mode at an intensity level of four. The initial cut was made in the upper central incisors and used as guides to determine the gingival zenith. Then, we cut the lateral incisors, canines, and upper first premolars. At the end of the procedure, the patient was instructed to consume soft and cold foods for at least 24 hours. We prescribed drug therapy with 500 mg acetaminophen every six hours for three days. Clinical follow-up was carried out at 48 hours and 15 days after surgery (figure 2A).

Figure 2 Images after the gingivectomy. A. Electric scalpel. B. 940 nm diode laser. C. 2780 nm Er,Cr:YSGG laser. 

940 nm diode laser gingivectomy: The area to be intervened was anesthetized with lidocaine spray. For the procedure, we used a BIOLASE® Epic X equipment, following a protocol of 1 W of power in a continuous mode with an E3-4 surgical tip (30 pm x 4 mm length), previously activated in articulating paper. At the end of the procedure, the patient was instructed to consume soft and cold food for at least 24 hours.

2780 nm Er,Cr:YSGG laser gingivectomy: Lidocaine spray was applied as topical anesthesia in the area to be operated. No infiltrative or regional anesthesia was used. We used a BIOLASE® Waterlase® equipment programmed at 2.5 W, 75Hz, H mode, air 20, water 40, and a gold sapphire tip MT4 (400 pm x 6 mm length). The procedure and tissue cutting were the same as those performed with the electric scalpel and diode laser (figure 2B and 2C).

In all cases, gingival tissue samples were taken and stored in 10% formaldehyde. Samples were sent to the histology laboratory for their examination using hematoxylin-eosin staining.

After the surgical procedures, follow-up was performed at eight days, observing satisfactory tissue response and absence of pain or sensitivity.

Histological report

The histological findings associated with the instrument’s effects were evaluated in the three histological sections. We measured the estimated depth of thermal damage, taking as a reference the basal membrane as the upper limit and the deepest point with collagen degeneration. Measures were made with the digital image processing software Imagen G®, version 1.50i.

Electric scalpel

Slide A (hematoxylin-eosin 10X): The section shows a mucous membrane where the squamous epithelium predominates. Although it can be recognized, it has undergone extensive changes due to burns and heat (enclosed in the oval), leaving a large part of carbonized tissue (figure 3A). The estimated depth of thermal damage was 200 pm.

Figure 3 Histological sections. A) Electric scalpel: ovals indicate extensive burn changes. B) Electric scalpel: intercellular edema (star), and the deep resection margin shows a clear border (black arrow) with no evidence of carbonization. C) 940 nm diode laser: basophilic changes due to collagen degeneration (square brackets). D) 2780 nm Er,Cr:YSGG laser: fine homogeneous basophilic band (red arrowhead) and signs of carbonization with sloughing (blue arrow) of the tissue. 

Slide B (hematoxylin-eosin 4X): This section shows gingiva with preserved mucosal epithelium, with some spongiotic changes - intercellular edema (star). The resection margin exposes a clear border (black arrow) with no evidence of carbonization or collagen degeneration (figure 3B).

940 nm diode laser (hematoxylin-eosin 4X): This section shows a gingiva fragment with preserved mucosal epithelium. Basophilic changes are evident due to collagen degeneration (in brackets), covering almost half of the tissue thickness in some areas. The estimated depth of thermal damage was 187 pm. The resection margin shows irregularity with a faint carbonization line (figure 3C).

2780 nm Er, Cr: YSGG laser (H-E 10x): The section reveals a deep resection margin with intense collagen degeneration extending only to the irregular connective tissue, evidenced by a thin homogeneous basophilic band (red arrowhead). The estimated depth of thermal damage was 27,2 pm. Fibroblasts preserved their morphology until the basophilic area. In epithelial tissue signs of carbonization are observed with tissue sloughing (blue arrow) and peripheral irregularity (figure 3D). Irregular dense connective tissue has abnormally dilated blood vessels. Regular dense connective tissue maintains its vascular network, fibroblasts have a fusiform appearance, and no degenerative changes are observed.

Ethical considerations

The present case follows Resolution 008430 of 1993, which establishes scientific, technical, and administrative standards for health research. The three patients were informed about the procedure to be conducted, the risks and benefits derived from it, and voluntarily signed a consent form for the performance of the surgical procedure and the donation of their samples for histological analysis.

Discussion

In dentistry, surgical cutting instruments have evolved. Currently, there are conventional scalpels, electric scalpels, and lasers, with varied applications in different dental specialties. All instruments incising the tissue generate lesions characteristic of its cutting mechanism and physical and biological properties 13. The cases presented in this paper confirm that the histological findings and the clinical characteristics of the wound vary according to the cutting equipment used. Although all instruments generate tissue lesions associated with thermal damage, when comparing them, we observed that some produce more damage than others.

Of the exposed cases, the 2780 nm Er,Cr:YSGG laser was the instrument that generated the least thermal damage. Cercadillo et al. confirmed it: The samples with less thermal effect were irradiated with Er,Cr:YSGG laser at 2780 nm and water/air spray, followed by the CO2 laser and 830 nm diode laser, concluding that the emission parameters of each system can impact the thermal damage inflicted on soft tissues. However, the wavelength of each laser influences tissue absorption rate and its reaction 14. This instrument superficial thermal effect in comparison with the electric scalpel and diode laser, lies in the spray of water attached to the cutting piece. According to its mechanism, the water is energized with the laser light, favors the cut, and serves as a cooling medium, contributing to tissue protection 15.

On the other hand, one of the samples taken with an electric scalpel revealed areas with tissue damage due to carbonization, while the second sample exposed cell preservation, and the thermal effect did not carbonize the tissue. This behavior suggests other variables influencing the tissue during cutting, such as the operator’s handling and the patient’s periodontal phenotype. A thick periodontal phenotype may imply more than one cut in the gingival margin to remove it completely. Therefore, the tissue will suffer more damage.

Regarding the effects of the electric scalpel, Ismail et al. indicated this incision produces thermal damage that includes carbonization, vacuolar degeneration, and nuclei elongation of the epithelial cells at the margins of the excision. Relevant findings with potential use in histopathological studies 16.

The 940 nm diode laser is a device with many indications in the dental field. Although it generated a thermal effect, we observed mucosa preservation and a poor carbonization area. Jin et al. compared wound healing after incisions with a conventional scalpel, a diode laser, and an Er,Cr:YSGG laser in the oral mucosa of guinea pigs and showed that the diode laser is a suitable device to use, but it produced more tissue damage than the scalpel or the Er,Cr:YSGG laser 17,18.

The clinical importance of these findings lies in the link between the histological condition and the postoperative period during the wound repair process 19. The epithelium healing route comprises a series of events in which the cellular and molecular dynamics of the injured tissue participate, seeking to restore integrity through epithelial migration that can be affected by different factors 20. During epithelium regeneration, the carbonization of the gingiva by the cutting instrument necessarily implies the necrosis of the burned area. The necrotic tissue acts as a barrier that interferes with the reparative action of the cells. Inflammation increases as leukocytes remove tissue debris through phagocytosis and lysis. In addition, necrotic tissue constitutes a niche for the proliferation of bacteria, increasing the risk of infection 21.

Some reported biological effects of the laser could improve or counteract the thermal damage generated by the irradiation of the gingival tissue during the cut. The laser light absorbed by the cells enhances ATP production, which induces cell migration and proliferation. On the other hand, the photobiomodulation effect of the laser promotes angiogenesis and local lymphatic drainage, leading to a decrease in inflammation and edema 22. These effects could be related to pain reduction, inflammation, and postoperative edema in patients treated with diode laser. Likewise, it presents an advantage over other instruments, such as the electric and the conventional scalpel.

Kaur et al. compared the efficacy and healing of soft tissue wounds using a diode laser (810 nm) versus the conventional scalpel approach in the second stage of implant surgery. They reported that laser use can minimize surgical trauma, reduce the amount of required anesthesia, improve visibility during surgery due to the absence of bleeding, and eliminate postoperative discomfort 23.

Conclusions

Under the parameters used in this study, the three cutting instruments generated histological changes produced by the thermal effect. The main ones were collagen coagulation and carbonization. According to our measurements, the Er,Cr:YSGG laser was the instrument that generated the least thermal damage to the tissue. Therefore, more studies are needed to confirm these results and allow a broad comparison between the three instruments.

Despite these findings, we found cell preservation indicators, implying other factors not related to the instrument that probably influence the generated effect on the tissue. These factors could be the periodontal characteristics of each patient and the operator’s skill during the procedure.

References

1. Storrer CL, de Oliveira ND, Deliberador TM, Ori LT, Guerrero SM, Santos FR, et al. Treatment of gingival smile: A case report. J Int Acad Periodontol. 2017;19:51-6. [ Links ]

2. Mostafa D. A successful management of sever gummy smile using gingivectomy and botulinum toxin injection: A case report. Int J Surg Case Rep. 2018;42:169-74. https://doi.org/10.1016/j.ijscr.2017.11.055Links ]

3. Kumar P, Rattan V, Rai S. Comparative evaluation of healing after gingivectomy with electrocautery and laser. J Oral Biol Craniofac Res. 2015;5:69-74. https://doi.org/10.1016/j.jobcr.2015.04.005Links ]

4. Kazakova RT, Tomov GT, Kissov CK, Vlahova AP, Zlatev SC, Bachurska SY. Histological gingival assessment after conventional and laser gingivectomy. Folia Med (Plovdiv). 2018;60:610-6. https://doi.org/10.2478/folmed-2018-0028Links ]

5. Kazakova R, Tomov G, Vlahova A, Zlatev S, Dimitrova M, Kazakov S, et al. Assessment of healing after diode laser gingivectomy prior to prosthetic procedures. Appl Sci (Basel). 2023;13:5527. https://doi.org/10.3390/app13095527Links ]

6. Mossaad AM, Abdelrahman MA, Kotb AM, Alolayan AB, Elsayed SA. Gummy smile management using diode laser gingivectomy versus botulinum toxin injection - a prospective study. Ann Maxillofac Surg. 2021;11:70-4. https://doi.org/10.4103/ams.ams_458_20Links ]

7. Donoso FA, Bizcar B, Sandoval C, Sandoval-Vidal P. Application of low power laser (LLLT) in pediatric patients: Review of the literature in a series of cases. Int J Odontostomat. 2018;12:269-73. https://doi.org/10.4067/S0718-381X2018000300269Links ]

8. Fekrazad R, Moharrami M, Chiniforush N. The esthetic crown lengthening by Er;Cr:YSGG laser: A case series. J Lasers Med Sci. 2018;9:283-7. https://doi.org/10.15171/jlms.2018.50Links ]

9. Nadhreen AA, Alamoudi NM, Elkhodary HM. Low-level laser therapy in dentistry: Extra-oral applications. Niger J Clin Pract. 2019;22:1313-8. https://doi.org/10.4103/njcp.njcp_53_19Links ]

10. Bhat P, Thakur SL, Kulkarni SS. Evaluation of soft tissue marginal stability achieved after excision with a conventional technique in comparison with laser excision: A pilot study. Indian J Dent Res. 2015;26:186-8. https://doi.org/10.4103/0970-9290.159159Links ]

11. Sobouti F, Rakhshan V, Chiniforush N, Khatami M. Effects of laser-assisted cosmetic smile lift gingivectomy on postoperative bleeding and pain in fixed orthodontic patients: a controlled clinical trial. Prog Orthod. 2014;15:66. https://doi.org/10.1186/s40510-014-0066-5Links ]

12. D’Arcangelo C, Di Nardo Di Maio F, Prosperi GD, Conte E, Baldi M, Caputi S. A preliminar study of healing of diode laser versus scalpel incisions in rat oral tissue: A comparison of clinical, histological, and immunohistochemical results. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007;103:764-73. https://doi.org/10.1016/j.tripleo.2006.08.002Links ]

13. Lione R, Pavoni C, Noviello A, Clementini M, Danesi C, Cozza P. Conventional versus laser gingivectomy in the management of gingival enlargement during orthodontic treatment: A randomized controlled trial. Eur J Orthod. 2020;42:78-85. https://doi.org/10.1093/ejo/cjz032Links ]

14. Cercadillo I, España A, Arnabat J, Valmaseda E, Berini L, Gay C. Histologic evaluation of thermal damage produced on soft tissues by CO2, Er,Cr:YSGG and diode lasers. Med Oral Patol Oral Cir Bucal. 2010;15:e912-8. https://doi.org/10.4317/medoral.15.e912Links ]

15. Abdelkarim H, Arnabat C, Parada I, Polonsky M, Arnabat J. Aerosols generation using Er,Cr:YSGG laser compared to rotary instruments in conservative dentistry: A preliminary study. J Clin Exp Dent. 2021;13:e30-6. https://doi.org/10.4317/jced.57731Links ]

16. Ismail A, Abushouk AI, Elmaraezy A, Menshawy A, Menshawy E, Ismail M, et al. Cutting electrocautery versus scalpel for surgical incisions: A systematic review and meta-analysis. J Surg Res. 2017;220:147-63. https://doi.org/10.1016/j.jss.2017.06.093Links ]

17. Jin JY, Lee SH, Yoon HJ. A comparative study of wound healing following incision with a scalpel, diode laser or Er,Cr:YSGG laser in guinea pig oral mucosa: A histological and immunohistochemical analysis. Acta Odontol Scand. 2010;68:232-8. https://doi.org/10.3109/00016357.2010.492356Links ]

18. Chandra GB, VinayKumar MB, Walavalkar NN, Vandana KL, Vardhan PK. Evaluation of surgical scalpel versus semiconductor diode laser techniques in the management of gingival melanin hyperpigmentation: A split-mouth randomized clinical comparative study. J Indian Soc Periodontol. 2020;24:47-53. https://doi.org/10.4103/jisp.jisp_186_19Links ]

19. Briceño JF, Gaviria DA, Carranza YA. Laser in Dentistry: Physical and biological foundations. Univ Odontol. 2016;35. https://doi.org/10.11144/Javeriana.uo35-75.loffLinks ]

20. Arango N, Betancur N, Gómez SH. Low-level laser effect on post-exodontia inflammation of the lower third molar. Rev Nac Odontol. 2018;13:1-10. https://doi.org/10.16925/od.v13i26.1640Links ]

21. Verma SK, Maheshwari S, Singh RK, Chaudhari PK. Laser in dentistry: An innovative tool in modern dental practice. Natl J Maxillofac Surg. 2012;3:124-32. https://doi.org/10.4103/0975-5950.111342Links ]

22. Garrigo MI, Valiente C. Biological effects of low power laser radiation on tissue repair. Rev Cubana Estomatol. 1996;33:60-3. [ Links ]

23. Kaur M, Sharma YPD, Singh P, Sharma S, Wahi A. Comparative evaluation of efficacy and soft tissue wound healing using diode laser (810 nm) versus conventional scalpel technique for second-stage implant surgery. J Indian Soc Periodontol. 2018;22:228-34. https://doi.org/10.4103/jisp.jisp_46_17Links ]

Citation: Orozco J, Rico D, Barrios L, Hoyos V, Blanco P. Histological artifacts associated with laser and electroscalpel gingivectomy: Case series. Biomédica. 2023;43:315-22. https://doi.org/10.7705/biomedica.6930

Author's contributions:

Jennifer Orozco: conceptualization, project management, data collection and analysis, content curation, research, writing - original draft, proofreading and editing

David Rico: conceptualization, data collection and analysis, provision of surgical instruments, writing - original draft

Lía Barrios: conceptualization, data collection and analysis, writing - original draft

Vivi Hoyos: data collection and analysis, content curation, writing - original draft, proofreading and editing

Pilar Blanco: conceptualization, data collection and analysis, provision of surgical instruments, revision and editing

Funding:

This research received no external funding.

Received: March 21, 2023; Accepted: August 22, 2023; Published: August 23, 2023

*Corresponding author: Jennifer Orozco, Corporación Universitaria Rafael Núñez, Carrera 17 N°35-117, Cartagena, Colombia Phone: (+57) (605) 643 9499, extensión 1007 jennifer.orozco@curnvirtual.edu.co

Conflicts of interest:

The authors declare no conflict of interest.

Creative Commons License This is an open-access article distributed under the terms of the Creative Commons Attribution License